The cookie settings on this website are set to 'allow all cookies' to give you the very best experience. Please click Accept Cookies to continue to use the site.

Mouse BAFF/CD257 (B-cell Activating Factor) CLIA Kit (MOES00004)

(No reviews yet) Write a Review
SKU:
MOES00004
$738

Description

Mouse BAFF/CD257 (B-cell Activating Factor) CLIA Kit

This kit uses Sandwich-CLIA as the method. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Mouse BAFF/CD257 . Standards or samples are added to the appropriate micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Mouse BAFF/CD257 and Avidin-Horseradish Peroxidase (HRP) conjugate are added to each micro plate well successively and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Mouse BAFF/CD257, biotinylated detection antibody and Avidin-HRP conjugate will appear fluorescence. The Relative light unit (RLU) value is measured spectrophotometrically by the Chemiluminescence immunoassay analyzer. The RLU value is positively associated with the concentration of Mouse BAFF/CD257. The concentration of Mouse BAFF/CD257 in the samples can be calculated by comparing the RLU of the samples to the standard curve.

Assay typeSandwich
Format96T
Assay time4.5h
ReactivityMouse
Detection methodChemiluminescence
Detection range15.63-1000 pg/mL
Sensitivity9.38 pg/mL
Sample volume100uL
Sample typeSerum, plasma and other biological fluids
RepeatabilityCV < 15%

Specificity

This kit recognizes Mouse BAFF/CD257 in samples. No significant cross-reactivity or interference between Mouse BAFF/CD257 and analogues was observed.

Typical Data

As the RLU values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

Concentration
(pg/mL)
RLU Average Corrected
100053483
53867
5367553650
50021782
23196
2248922464
25011097
9325
1021110186
1254767
5035
49014876
62.52534
2372
24532428
31.251345
1217
12811256
15.63704
712
708683
024
26
25--

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level Mouse BAFF/CD257 were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level Mouse BAFF/CD257 were tested on 3 different plates, 20 replicates in each plate.

  Intra-assay Precision Inter-assay Precision
Sample123123
n202020202020
Mean (pg/mL)48.03102.23393.1346.1796.36380.24
Standard deviation5.248.2630.114.9811.4123.77
C V (%)10.918.087.6610.7911.846.25

Recovery

The recovery of Mouse BAFF/CD257 spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range (%) Average Recovery (%)
Serum (n=5)94-107101
EDTA plasma (n=5)93-10398
Cell culture media (n=5)89-10195

Linearity

Samples were spiked with high concentrations of Mouse BAFF/CD257 and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

    Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2Range (%)94-10888-10588-98
Average (%)1029693
1:4Range (%)85-9799-11487-100
Average (%)9210593
1:8Range (%)88-10298-11491-105
Average (%)9410696
1:16Range (%)90-103102-11994-107
Average (%)97110100

Kit Components & Storage

An unopened kit can be stored at 4'C for 1 month. If the kit is not used within 1 month, store the items separately according to the following conditions once the kit is received.

Item Specifications Storage
Micro CLIA Plate(Dismountable)8 wells ×12 strips-20'C, 6 months
Reference Standard2 vials
Concentrated Biotinylated Detection Ab (100×)1 vial, 120 uL
Concentrated HRP Conjugate (100×)1 vial, 120 uL-20'C(shading light), 6 months
Reference Standard & Sample Diluent1 vial, 20 mL4'C, 6 months
Biotinylated Detection Ab Diluent1 vial, 14 mL
HRP Conjugate Diluent1 vial, 14 mL
Concentrated Wash Buffer (25×)1 vial, 30 mL
Substrate Reagent A1 vial, 5 mL4'C (shading light)
Substrate Reagent B1 vial, 5 mL4'C (shading light)
Plate Sealer5 pieces
Product Description1 copy
Certificate of Analysis1 copy
View AllClose

Additional Information

Reactivity:
Mouse
ELISA Type:
CLIA
View AllClose